100 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System
9
5. To change this number, press the SELECT
button, and then press the UP or DOWN
button to select your high glucose
alert level.
Your high glucose alert value can be
set between 120 and 400 mg/dL in 10
mg/dL steps.
Your low glucose alert value can be
set between 60 and 100 mg/dL in
5 mg/dL steps.
6. Press the SELECTbuttontoconrmyouralertlevel.
7. Press the LEFTbuttontoreturntotheAlertsmenuwhenyounish.
9.2 ADVANCED ALERTS
Advanced alerts include the high and low snooze, rise and fall rate, and
out of range alerts. All advanced alerts follow these steps.
1. Press the SELECT button to turn on the receiver.
2. Press the SELECT button to enter the Main Menu.
3. Press the UP or DOWN button to highlight “Alerts,” and press the
SELECT button.
4. From the Alerts menu, press the UP or
DOWN button to select “Advanced,” and
press the SELECT button.
Alerts menu,
Advanced highlighted
High Alert setting screen,
200 mg/dL selected